Microtiter Plate Reader

Advanced Instruments LLC Artel Artel MVS 800 BART Microtiter Plate Reader

Request a Quote

Fill in your details below to receive a quote on this item. Our team will be in touch as soon as possible.

Note: We require email addresses from workplace domains only. If you have an email address containing gmail.com, yahoo.com, etc. please call 1-877-267-4272 to talk to a representative.

Model No: 23272

Serial No: 203824

Product ID: ALE-12595

System Capabilities:

• Measures performance over a wide volume range of 0.0001 to 350 μL for Automated Liquid Handlers and

Multichannel Pipettes

• Measures aqueous and non-aqueous volume transfers, including dimethyl sulfoxide (DMSO)) and PCRMix

• Obtains volume statistics tip-by-tip and well-by-well

• Eliminates the need for rigorous environmental controls

• Simplifies verification and calibration—easy-to-use for technicians of any skill level

• Generates measurement results traceable to the International System of Units (SI) via standards

maintained by NIST.

Key Applications:

• Optimizing liquid classes

• Facilitating of scale-up and method transfer

• Developing, troubleshooting, and validating assays, methods, and SOPs

• Ensuring reproducibility

• Site acceptance testing and establishing baseline performance of new equipment

• Verifying performance before and after maintenance

• Assessing performance of other essential liquid handling equipment such as

plate shakers, plate washers, and bulk dispensers

• Assessment and training of multichannel pipette operators

Compliance:

The MVS complies with the Ratiometric Photometric Method according

to ISO/IWA 15:2015 and can be used to fulfill test and calibration

requirements of handheld or automated liquid handling systems according

to CLSI QMS23:2019, ISO 17025:2017, cGMP, and cGLP. Data Manager

Software can be used in laboratories compliant to 21 CFR Part 11


ARTEL 800TSNB PLATE READER

Dimensions (DxWxH) 41.9 x 38.1 x 17.8 cm

Weight 9.97 kg

Display Touch screen

Light source Tungsten gas filled bulb

Wavelength selection Metal oxide interference filters

Center wavelengths:

520.2 nm (bandwidth at half-max = 6.2 nm)

730.5 nm (bandwidth at half-max = 10 nm)

Additional filters 405 nm, 450 nm, 490 nm

COM port USB

Power requirements Voltage: 90 – 260 VAC Frequency: 50 – 60 Hz Current: 2 A maximum

MVS TITER PLATE SHAKER

Dimensions (LxWxH) 142 x 99 x 48.2 mm

Amplitude 2.0 mm orbital

Shaking speed range 200 to 3000 rpm

Power requirements External power supply - 100-240 V AC, 50-60 Hz

MVS CALIBRATOR PLATE

Storage Temperature 15 - 25 °C in protective case, out of direct light

Operation Temperature 15 - 30 °C

Recalibration Frequency 12 month factory recalibration required

Shelf life 1 year

MVS BAR CODE READER

Dimensions (LxWxH) 104 x 71 x 160 mm

Weight 147 g

MVS COMPUTER (minimum requirements)

Processor x64 Processor: 1.4 GHz or faster

Memory 1 GB

I/O Ports 3 USB Connectors

Network 10/100 Ethernet and 802.11b Wireless

Storage 20 GB of hard drive space

Supported Operating Systems Windows® 7 (64-bit) or Windows® 10 (64-bit)